    Magnetic stirring electric heat set does not stir how to return a responsibility?
    Date of publication:2019-07-11

    Magnetic stirring electric heat set does not stir how to return a responsibility?

    There are many reasons. First, check whether the power supply is electrified. Second, check whether the fuse on the equipment is blown.

    Here are some tips for using magnetic heating agitator:

    1. Do not dry burn.

    2. The instrument should be well grounded.

    3. When ** is used for the second time, white smoke and peculiar smell come out of the board, which is a normal phenomenon. Due to the oil and other compounds contained in the production process of the internal insulation material, it should be placed in a ventilated place and used normally after several minutes of disappearance.

    4. Since the heat resistance of the motor and magnet of the agitator is limited, the instrument cannot be used only for heating in the heating test, especially in the high temperature test. Please adjust the motor to the rotating or low-speed rotating state (idling) to prevent the motor, magnet and high temperature radiation from being damaged.

    5. At the end of high temperature heating, please turn off the heating and leave the remaining heat for a few minutes before turning it off and stirring.

    6. The heating part has a high temperature, so be careful to avoid scalding.

    7. If there is wet hand, liquid overflow or long-term exposure to excessive humidity, there may be inductive electricity transmitted to the shell through the insulation, please keep the grounding wire and pay attention to ventilation. In case of serious electric leakage, please do not use it again. Please put it in the sun to dry or put it in the oven to dry before using, so as to avoid danger.

    8. If no power is found, please check whether the right rear insurance needs to be replaced.

    9. When not in use for a long time, please keep the instrument clean and keep it in a dry and non-corrosive place.