- Digital display constant temperature magnetic agitator supply
- Digital display constant temperature magnetic agitator supply
The digital display constant temperature magnetic agitator completely enclosed heating plate can be used as auxiliary heating, the temperature can be set and displayed, the digital display is intuitive and accurate. Agitator made of ptfe and high quality magnetic steel, resistant to high temperature, wear resistance, chemical corrosion resistance and strong magnetism. It can conduct precise and stable constant temperature agitation of liquid solution in a wide range of speed and temperature, especially suitable for stirring small samples. Digital display constant temperature magnetic agitator can be widely used in petroleum, chemical industry, medicine and health, environmental protection, biochemistry, education and scientific research, etc.
Operation method of constant temperature magnetic agitator with digital display:
1. Place the magnetic stirring rod in a beaker containing the solution. Place the beaker on the heating plate and insert the sensor element.
2. Turn on the power, adjust the heating speed and start stirring.
3. When stirring, adjust the speed control button slowly. Too fast adjustment will make the stirring rotor jump out of the magnetic steel force. Should quickly turn the knob to the stop position, wait for the agitator to be stationary, slowly rise the speed of stirring, gradually stable rise the speed. At room temperature, the viscosity of the liquid, often poor thermal conductivity, heating stirring, should not be rapidly warming, lest the vessel rupture. Take full advantage of the thermostat, gradually step by step temperature rise, and sensor elements must be inserted into the outer water jacket.
4. Predict the temperature in the container by slowly turning the thermostat to lower the temperature indicating red label. When the red label is on, the temperature indicating red label is the temperature of liquid inserted into the measuring element.
Notes for use of digital display thermostatic magnetic agitator:
1. The device must be reliably grounded to ensure the safety of equipment and person.
2. When stirring, adjust the speed control button slowly. Too fast adjustment will make the stirring rotor jump out of the magnetic steel force. Should quickly turn the knob to the stop position, wait for the agitator to be stationary, slowly rise the speed of stirring, gradually stable rise the speed.
3. For liquids with high viscosity at room temperature, their thermal conductivity is often poor (such as epoxy resin). When heating and stirring, it is not appropriate to heat up rapidly to avoid vessel rupture. Take full advantage of the thermostat, gradually step by step temperature rise, and sensor elements must be inserted into the outer water jacket.
4. If the aluminum plate on the surface of the heating plate is covered with liquid, it will corrode the plate surface or generate heat and steam, which will affect the heating element and the motor. Please turn off the power immediately to remove it.